Even After Cleaning, Microbes Linger on Endoscopes

This article originally appeared here.
Even After Cleaning, Microbes Linger on Endoscopes
Even After Cleaning, Microbes Linger on Endoscopes

(HealthDay News) — Despite adherence to current U.S. reprocessing guidelines, microbes and biologic debris persist on endoscopes, according to research published in the August 1 issue of the American Journal of Infection Control.

Cori Ofstead, MSPH, of Ofstead & Associates in St. Paul, MN, and colleagues examined 15 colonoscopes and gastroscopes used in gastrointestinal procedures after they underwent each step of recommended cleaning guidelines. Those steps included: bedside cleaning; manual cleaning in dedicated reprocessing rooms; and automated cleaning with a high-level disinfectant. The scopes were stored vertically after drying with isopropyl alcohol and forced air. If contamination levels exceeded pre-determined targets for each cleaning step, the cleaning procedure was repeated and the scope retested.

Surviving microbes were found on 92% of the devices after bedside cleaning, 46% after manual cleaning, 64% after high-level disinfection, and 9% after overnight storage, the researchers found. Residual contamination above target levels was found on 100% of the devices after bedside cleaning, 92% after manual cleaning, 73% after high-level disinfection, and 82% after overnight storage.

The researchers do hold out hope that other methods might get hospitals closer to complete disinfection, however. "More research is needed to identify processes that can ensure all flexible endoscopes are free of residual contamination and viable microbes prior to patient use," the authors said in a news release from the Association for Professionals in Infection Control and Epidemiology. "Results from this study suggest that current standards and practices may not be sufficient for detecting and removing residual contamination."

3M provided funding and materials for the study.